IF: 0.644
REUTERS THOMSON

Comparison of Rectal Dose in Different Techniques of Prostate Cancer External-Beam Radiotherapy Based on TLD and XR Type T GAFCHROMIC® Film Dosimetry

AUTHORS

F Ebrahimi-Tazehmahalleh 1 , * , H Gholamhosseinian 2 , M Layegh 3 , N Ebrahimi-Tazehmahalleh 4 , H Esmaily 5

AUTHORS INFORMATION

1 Department of Medical Physics, Faculty of Medicine, Mashhad University of Medical Sci-ences, Mashhad, Khorasan Razavi, Iran

2 Department of Radiotherapy, Imam Reza Hospital, Mashhad University of Medical Science, Mashhad, Khorasan Razavi, Iran

3 Motahari Clinic, Khoy, West Azarbaijan, Iran

4 Department of BioStatistics, Faculty of Medicine, Mashhad Univer-sity of Medical Sciences, Mashhad, Khorasan Razavi, Iran

5 Department of BioStatistics, Faculty of Medicine, Mashhad University of Medical Sci-ences, Mashhad, Khorasan Razavi, Iran

How to Cite: Ebrahimi-Tazehmahalleh F, Gholamhosseinian H, Layegh M, Ebrahimi-Tazehmahalleh N, Esmaily H. Comparison of Rectal Dose in Different Techniques of Prostate Cancer External-Beam Radiotherapy Based on TLD and XR Type T GAFCHROMIC® Film Dosimetry, Iran Red Crescent Med J. 2008 ; 10(4):e95871.

ARTICLE INFORMATION

Iranian Red Crescent Medical Journal: 10 (4); e95871
Published Online: October 31, 2008
Article Type: Research Article
Received: July 01, 2019
Accepted: October 31, 2008

Crossmark

CHEKING

READ FULL TEXT
Abstract

Background: Radiotherapy of prostate carcinoma often results in high doses to surrounding structures, such as the rectum and bladder. Therefore, these organs should be closely monitored. The aim of this study is to evaluate the dose received by the target volume and rectum to compare two different methods of dose measurement with each other and to check the homogeneity of dose in the tumor volume.

 

Methods: The dose distribution throughout a planned target volume and the rectum (OaR) in a phantom exposed to 9 MV photon beam, similar to treatment conditions were studied. Several techniques of external beam radiation therapy such as two-, three- and four-field have been planned. Dosimetry was performed using GAFCHROMIC® film and TLD-100 chips.

 

Results: The rectal and cancer volume measured doses in treatment were similar to the prescribed doses. The results of two dosimetry types were compared with each other as well as with treatment planning. Rectal dose in three- and four-field (equal tumor dose and equal applied dose) techniques were respectively 23.15, 28.87 and 15.22% lower than the tumor dose.

 

Conclusion: There was not a statistically significant difference between received and prescribed doses. So, this study showed that the Gafchromic film dosimetry can be used for fast dosimetric evaluations.

Keywords

Rectal dose Radiation therapy TLD XR type T film Prostate cancer

© 2008, Author(s). This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/) which permits copy and redistribute the material just in noncommercial usages, provided the original work is properly cited.