IF: 0.644
REUTERS THOMSON

Drug-Eluting Stent Fracture within 6 Days after Stent Implantation

AUTHORS

J Zamany 1 , Y Mahmoody 2 , * , MA Ostavan 1

AUTHORS INFORMATION

1 Cardiovascular Research Center, Faghihi Hospital, Shiraz University of Medical Sciences, Fars, Iran

2 Cardiovascular Research Center, Faghihi Hospital, Shiraz University of Medical Sciences, mahmoody_6@yahoo.com, Fars, Iran

How to Cite: Zamany J, Mahmoody Y, Ostavan M. Drug-Eluting Stent Fracture within 6 Days after Stent Implantation, Iran Red Crescent Med J. Online ahead of Print ; 13(2):139-140.

ARTICLE INFORMATION

Iranian Red Crescent Medical Journal: 13 (2); 139-140
Article Type: Case Report
Received: May 10, 2010
Accepted: August 2, 2010

Crossmark

CHEKING

READ FULL TEXT
Abstract

A 63-year-old woman with typical chest pain referred to our center. Her risk factors were hypertension (HTN), hyperlipidemia (HLP), diabetic mellitus (DM), and smoking. Coronary angiography (CAG) revealed a tubular lesion (95%) in proximal part, and a diffuse lesion (95%) in the mid part of left anterior descending (LAD) with a good distal flow. Stenting of proximal and mid part lesion was done for her with 2 drug-eluting stent (paclitaxel-eluting coronary stent system, EURoCOR) with success. After 6 days, due to a typical chest pain without response to medications, the coronary angiography was performed showing a definite fracture of stent which was implanted in the mid part of left anterior descending artery (LAD) 6 days before admission.  

Keywords

Drug-eluting stent Fracture Implantation

© 0, Author(s). This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/) which permits copy and redistribute the material just in noncommercial usages, provided the original work is properly cited.